3.2 Remote control commands

758 KFD Titrino

117

Reports which are sent from the Titrino are marked with space (ASCII 32) and

' at the beginning. Then an individual identifier for each report follows. Reports

which are triggered by RS232 ($G) have the same introducer but without pre-

ceding space, i.e. they start with '.

3.2.2.102. Info.CalibrationData $G

Info.CalibrationData.Inp1.pHas –20.00...7.00...+20.00

Info.CalibrationData.Inp1.Slope –9.999...1.000...+9.999

Info.CalibrationData.Inp1.Temp –170.0...25.0...+500.0

Info.CalibrationData.Inp1.Date read

only

Info.CalibrationData.Inp1.ElectrodeId read

only

identical for .Inp2 and .Diff

pH calibration data for measuring input 1. After the calibration, the data are

entered automatically together with the date of the calibration and the elec-

trode identification, see 3.2.2.42.

Calibration data can be entered. They are accepted with &Info.CalibrationData

$G. If calibration data are entered, the calibration date is deleted.

3.2.2.103. Info.PrepData.D0.Date read

only

Info.PrepData.D0.Time read

only

Info.PrepData.D1.Type read

only

Info.PrepData.D1.Date read

only

Info.PrepData.D1.Time read

only

identical to .D2

Data from the last correctly executed preparation.

3.2.2.104. Info.Checksums $G

Info.Checksums.MPList read

only

Info.Checksums.ActualMethod read

only

The checksums can be used to identify the content of a file unequivocally,

e.g. files with identical content

have identical results of the checksums. An empty file has checksum "0". The

calculation of the checksums is triggered with $G.

.MPList: Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the

same results of the checksum.

3.2.2.105. Info.DetermData $G

Info.DetermData.Write ON,

OFF

Info.DetermData.MPList.1.Attribute read

only/read+write

Info.DetermData.ExV read

only/read+write

Info.DetermData.MPList.1.X read

only/read+write

Info.DetermData.MPList.1.Y read

only/read+write

Info.DetermData.MPList.1.Z1 read

only/read+write

Info.DetermData.MPList.1.Z2 read

only/read+write

for every measuring point

Determination data in hexadecimal format. A measuring point list is available

in mode SET, KFT and MEAS.