Metrohm 758 KFD Titrino Manuel d'utilisation

Page 106

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

102

3.2.2.16. Mode.Parameter.TitrPara.XPause

0...999999

Parameter for SET, KFT: Pause time in s. Runs before dosing the start vol-

ume.

3.2.2.17. Mode.Parameter.TitrPara.StartV.Type abs.,

rel.,

OFF

Mode.Parameter.TitrPara.StartV.V 0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999999

Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.

Parameters for SET, KFT: Start volume.

If an absolute start volume (abs.) has been selected, the volume in mL is

valid.

A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size ∗ factor

The factor is valid.

The dispensing rate in mL/min applies to both cases. Max. means maximum

possible dispensing rate with the Exchange Unit in current use.

3.2.2.18. Mode.Parameter.TitrPara.Pause

0...999999

Parameters for SET, KFT: Pause time in s. Is waited off after the dispensing of

the start volume.

3.2.2.19. Mode.Parameter.TitrPara.ExtrT

0...999999

Parameter for SET, KFT: Extraction time in s.

3.2.2.20. Mode.Parameter.TitrPara.DosUnit internal

D0, external D1/D2

Mode.Parameter.DosPara.DosUnit internal

D0, external D1/D2

.TitrPara.DosUnit: Parameter for SET, KFT. Selection of dosing unit.

3.2.2.21. Mode.Parameter.TitrPara.MeasInput

1, 2, diff.

Mode.Parameter.TitrPara.Ipol –127...1...+127

Mode.Parameter.TitrPara.Upol –1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest ON, OFF

Parameters for SET, KFT:

Selection of the measuring input; valid with measured quantities pH and U.

"diff." means differential amplifier, see page 167.
With Ipol, the inquiries for the polarization current in µA (Ipol) and

.PolElectrTest are valid. Standard for Ipol with KFT is 50 µA.

With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in

steps of 10 mV.

Besides .PolElectrTest is valid.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to an active state (titration or conditioning).