Metrohm 758 KFD Titrino Manuel d'utilisation

Page 117

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

113

3.2.2.79. Config.RSSet1 $G

Config.RSSet1.Baud

300, 600, 1200, 2400, 4800, 9600,

19200, 38400, 57600, 115200

Config.RSSet1.DataBit 7,

8

Config.RSSet1.StopBit

1, 2

Config.RSSet1.Parity even,

odd, none

Config.RSSet1.Handsh HWs, SWchar, SWline, none

$G sets all RS settings. The changes are performed only if the instrument is

inactive. After the setting of the interface parameters, wait at least 2 s to allow

the components to equilibrate.

Settings of the values for the data transmission via the RS interface: baud

rate, data bit, stop bit, parity and type of handshake, see also page 130ff.

Baud rates >9600 need a PC which is equipped accordingly (e.g. with 16550

component).

The setting of the values must be initiated with $G immediately after entry of

the values.

3.2.2.80. Config.ComVar.C30

with up to .C39, etc.

0... ±999999

Values of the common variables from C30 up to C39. Insert the common

variables directly or describe the determination results directly from the

method, see 3.2.2.7.

3.2.2.81. Config.DosPrep.PowerOnPrep ON,

OFF

Warning after power on, that the preparation of the buret has to be executed.

3.2.2.82. Config.DosPrep.Report ON,

OFF

Report output after preparation of burets.

3.2.2.83. Config.DosPrep.Select internal

D0, external D1, external D2

Selection of buret.

3.2.2.84. Config.DosPrep.DX.WarnInterval 5...9999,

OFF

Time interval in min for automatic warning that a preparation has to be exe-

cuted.

3.2.2.85. Config.DosPrep.DX(.Dosimat).V 0...3.5...99999.99

Volume in mL, which will be expelled at the preparation. Start sequence see

3.2.2.121.

3.2.2.86. Config.DosPrep.DX(.Dosimat).Repeat 1...2...9

Number of cycles for expelling the volume at the preparation. Start sequence

see 3.2.2.121.

Advertising