Metrohm 758 KFD Titrino Manuel d'utilisation

Page 131

Advertising
background image

3.2 Remote control commands

758 KFD Titrino

127

.CyclNo: Cycle number. Together with the cycle time (see 3.2.2.118), a

time frame can be set up. The other data belong to the corre-

sponding cycle number. The cycle number is set to 0 at the start

of a method and it is incremented until the end of the method.

.V: Volume.

.dVdt:

associated volume drift.

.dMeasdt: associated measured value drift.

.dMeasdV: associated 1st derivative of the titration curve.

The unit "titrator" must be preset (see 3.2.2.136).

3.2.2.138. Setup.AutoInfo.Status ON, OFF

Setup.AutoInfo.P ON, OFF

Setup.AutoInfo.T.R ON, OFF

Setup.AutoInfo.T.G ON, OFF

Setup.AutoInfo.T.GC ON, OFF

Setup.AutoInfo.T.S ON, OFF

Setup.AutoInfo.T.B ON, OFF

Setup.AutoInfo.T.F ON, OFF

Setup.AutoInfo.T.E ON, OFF

Setup.AutoInfo.T.H ON, OFF

Setup.AutoInfo.T.C ON, OFF

Setup.AutoInfo.T.O ON, OFF

Setup.AutoInfo.T.N ON, OFF

Setup.AutoInfo.T.Re ON, OFF

Setup.AutoInfo.T.Si ON, OFF

Setup.AutoInfo.T.M ON, OFF

Setup.AutoInfo.T.EP ON, OFF

Setup.AutoInfo.T.RC ON, OFF

Setup.AutoInfo.C.B1 ON, OFF

Setup.AutoInfo.C.R1 ON, OFF

Setup.AutoInfo.C.B2 ON, OFF

Setup.AutoInfo.C.R2 ON, OFF

Setup.AutoInfo.I ON, OFF

Setup.AutoInfo.O ON, OFF

ON means that the Titrino reports automatically the moment the correspond-

ing change occurs.

.Status:

Global switch for all set AutoInfo.

.P

PowerOn: Simulation of power on (3.2.2.140). Not from mains.

Messages from node .T, Titrator:

.T.R

Ready: Status 'Ready' has been reached.

.T.G

Go: Instrument has been started.

.T.GC

GoCommand: Instrument (or submethod in TIP) has received a go

command.

.T.S

Stop: Status 'Stop' has been reached.

.T.B

Begin of sequence (or submethod).

.T.F

Final: End of determination (or submethod), the final steps will be

carried out.

.T.E

Error. Message together with error number, see page 76ff.

.T.H

Hold: Status 'Hold' has been reached.

.T.C

Continue: Continue after hold.

.T.O

Conditioning OK: EP reached (in SET, KFT with conditioning).

Advertising