3.2 Remote control commands

758 KFD Titrino

114

3.2.2.87. Config.DosPrep.DX(.Dosimat).DosRate 0.01...150,

max.

Config.DosPrep.DX(.Dosimat).FillRate 0.01...150,

max.

Dosing and filling rate in mL/in for the preparation. Start sequence see

3.2.2.121.

3.2.2.88. Config.DosPrep.DX.Select Dosimat, Dosino

Selection of buret type. Start sequence see 3.2.2.121.

3.2.2.89. Config.DosPrep.DX.Dosino.Outlet

tip, flask

Place where the liquid of the preparation is expelled. Start sequence see

3.2.2.121.

tip: To the tip.

flask: Back to the flask.

3.2.2.90. Config.DosPrep.DX.Dosino.DosTubing.Length 1...40...999.9

Config.DosPrep.DX.Dosino.DosTubing.Diameter 0.1...2...9.9

Config.DosPrep.DX.Dosino.AspirTubing.Length 1...25...999.9

Config.DosPrep.DX.Dosino.AspirTubing.Diameter 0.1...2...9.9

.DosTubing: Dosing tube. Length in cm, diameter in mm.

.AspirTubing: Aspiration tube. Length in cm, diameter in mm.

Start sequence see 3.2.2.121.

3.2.2.91. Config.DosPrep.DX.Dosino.DosRate 0.01...150,

max.

Config.DosPrep.DX.Dosino.FillRate 0.01...150,

max.

Dosing and filling rate in mL/min for preparation. Start sequence see

3.2.2.121.

3.2.2.92. SmplData.Status ON, OFF

On/off switching of silo memory. When the silo memory is switched on, the

sample data are fetched from the lowest valid silo line.

3.2.2.93. SmplData.OFFSilo.Id1

up to 8 ASCII characters

SmplData.OFFSilo.Id2

up to 8 ASCII characters

SmplData.OFFSilo.Id3

up to 8 ASCII characters

SmplData.OFFSilo.ValSmpl

6-digits, sign and decimal point

SmplData.OFFSilo.UnitSmpl

up to 5 ASCII characters

Current sample data.

The identifications Id1...Id3 can be used in formulas as sample-specific calcu-

lation constants C21...C23.

If "no unit" is desired for the unit of the sample size, the blank string must be

entered.