Capacity card, Closing and setup protection – Ohaus I5S Indicator Manuel d'utilisation
Page 9

8
Scale Base Wire Connections
Bottom View
PC Board
+ EXCITATION
+ SIGNAL
- EXCITATION
- SIGNAL
GROUND
4.
Connect the scale base wires to the
terminal block located on the bot-
tom of the printed circuit board as
shown in the illustration.
Lock-out
Switch
Battery Cable
Connector
Slot for Battery
Cable
Top View
PC Board
5.
Reconnect the battery connector to
the circuit board.
6.
Push the circuit board back into the
housing.
Capacity Card
Capacity Card
Capacity Card
Capacity Card
Capacity Card
Type or write the capacity and readabil-
ity data for the scale base on one of the
perforated cards and slide it under the
front panel label near the display.
Closing and Setup Protection
Closing and Setup Protection
Closing and Setup Protection
Closing and Setup Protection
Closing and Setup Protection
•
If the setup information (explained in
Chapter 2) does not need to be pro-
tected from unauthorized changes,
position the battery cable coming from
the end cap into the slot on the end of
the PC board, then replace the end
cap.
•
If you intend to lock out changes to the
setup selections you make, leave the
end cap in position but do not tighten
the screws. You will need to access
the menu lockout switch located on
the circuit board following the setup
procedure.
•
If the unit is being powered by batter-
ies during setup rather than the AC
adapter, and you want to lock out
setup changes, replace the end cap
now and remove it after setup to set
the menu lock-out switch.
Off
On
1
1
1
1
1
Introduction and Installation