Zero – Ohaus I5S Indicator Manuel d'utilisation
Page 16

15
Zero
Zero
Zero
Zero
Zero
Zero specifies the percentage of full
capacity load (2% or 100%) that may be
cleared by pressing
.
☞
A 2% Zero setting is used for hop-
per scales or other large scales
where an accidental zero would
lose the current weight. A 2% Zero
is also required for Canadian Le-
gal for Trade applications. Bench
and counter scales normally use
100% Zero setting.
To access the current setting, press
when ZERO is displayed.
1.
To change the displayed setting,
press
.
2.
To accept the displayed setting,
press
.
To proceed to the next parameter (Cali-
bration Unit), press
when ZERO is
displayed.
To exit the Setup submenu and store
your settings, repeatedly press
until END is displayed, then press
to return to the main menu.
To exit main menu and return to a weigh-
ing mode, repeatedly press
until
END is displayed then press
. The
scale is now in a weighing mode.