Auto Zero Tracking
Auto Zero Tracking
Auto Zero Tracking
Auto Zero Tracking
Auto Zero Tracking
Auto Zero Tracking compensates for
changes in zero reading due to temper-
ature changes or build up of small
amounts of debris on the scale base
platform. Select (0.5, 1, or 3 divisions or
OFF), the value of weight change that
the indicator will disregard in order to
maintain a zero reading.
Auto Zero Tracking occurs only when
the display indicates zero regardless of
the load on the scale base.
NOTE
NOTE
NOTE
NOTE
NOTE: If LFT is set ON, the Auto-Zero
Tracking OFF setting is not available.
To access the current setting, press
when Auto-0 is displayed.
1.
To change the displayed setting,
press
.
2.
To accept the displayed setting,
press
.
To proceed to the next parameter (Auto
Off Timer), press
when Auto-0 is
displayed.
To exit the Auto-Zero Tracking submenu
and store your settings, repeatedly press
until END is displayed, then press
to return to the main menu.
To exit main menu and return to a weigh-
ing mode, repeatedly press
until
END is displayed then press
. The
scale is now in a weighing mode.