Setup submenu, Legal for trade – Ohaus I5S Indicator Manuel d'utilisation
Page 15

14
Setup Submenu
Setup Submenu
Setup Submenu
Setup Submenu
Setup Submenu
1.
With the indicator OFF, press and
hold
until SETUP is dis-
played, then release it.
2.
When SETUP is displayed, press
to enter the submenu. The
first parameter, Legal For Trade, will
be displayed.
Legal For Trade (LFT)
Legal For Trade (LFT)
Legal For Trade (LFT)
Legal For Trade (LFT)
Legal For Trade (LFT)
This parameter is used to configure the
I
5S for Legal for Trade or non Legal for
Trade use. Setting this parameter to ON
automatically sets certain parameters for
legal for trade requirements, disables
others, and limits overall display resolu-
tion to 1:5,000.
To access the current setting (ON or
OFF), press
when LFT is dis-
played.
1.
To change the displayed setting,
press
.
2.
To accept the displayed setting,
press
. To proceed to the
next parameter (Zero), press
when LFT is displayed.
To exit the Setup submenu and store
your settings, repeatedly press
until END is displayed, then press
to return to the main menu. To exit main
menu and return to a weighing mode,
repeatedly press
until END is dis-
played then press
. The scale is
now in a weighing mode.
Legal for Trade OFF
Zero
100%
Calibration
Unit lb
Full Scale
Capacity 5000
Graduation Size 1
Calibration Point 5000
Menu Lock-out Switch
Setup ON
Scale ON
Calibration ON
RS232 ON
SETUP MENU
SETUP MENU
SETUP MENU
SETUP MENU
SETUP MENU
FACTORY DEFAULTS
FACTORY DEFAULTS
FACTORY DEFAULTS
FACTORY DEFAULTS
FACTORY DEFAULTS
2
2
2
2
2
Setup and Calibration