Graduation size – Ohaus I5S Indicator Manuel d'utilisation
Page 19

18
Graduation Size
Graduation Size
Graduation Size
Graduation Size
Graduation Size
This parameter specifies the increments
in which the display will count.
If Legal For Trade is set to ON, resolu-
tion (Capacity
÷
Graduations) is limited
to a maximum of 5000 divisions. The
graduation selections available are lim-
ited by the capacity entered under Full
Scale Capacity to prevent a combination
which exceeds 5,000 divisions.
If the graduation size selected does not
meet the resolution requirement, the
Setup menu cannot be exited. An error
message is displayed and you are auto-
matically returned to the GRAD param-
eter. A valid graduation value must be
selected.
To access the current setting, press
when GRAD is displayed.
1.
To change the displayed setting,
press
.
2.
To accept the displayed setting,
press
.
To proceed to the next parameter (Cali-
bration Point), press
when GRAD
is displayed.
To exit the Setup submenu and store
your settings, repeatedly press
until END is displayed, then press
to return to the main menu.
To exit main menu and return to a weigh-
ing mode, repeatedly press
until
END is displayed then press
. The
scale is now in a weighing mode.
Complete List of Graduation
Complete List of Graduation
Complete List of Graduation
Complete List of Graduation
Complete List of Graduation
Selections
Selections
Selections
Selections
Selections
.0005, .001, .002,
.005, .01, .02, .05,
.1, .2, .5, 1
1
1
1
1, 2, 5, 10
2
2
2
2
2
Setup and Calibration