Chapter 5: Using the Protector Perchloric Acid

Product Service 1-800-522-7658

33

Original instructions

ductwork. This is done by connecting the water line to the plugged connection on
top of the washdown hood system or ordering additional washdown valves found
in Chapter 2 and Chapter 7.

Working in Your Protector Perchloric Acid

Fume Hood

Planning

Thoroughly understand procedures and equipment required before

beginning work.

Arrange for minimal disruptions, such as room traffic or entry into the

room while the hood is in use.

Start-up

Turn on fluorescent light and hood blower.

Slowly raise the sash.

Check the baffle air slots for obstructions.

Allow the hood to operate unobstructed for 5 minutes.

Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.

Wear a protective mask if appropriate.

Loading Materials and Equipment

Only load the Perchloric Acid Hood materials required for the procedure.

Do not overload the hood. Only use Perchloric acid in the Perchloric Acid
Fume Hood.

Radioisotope materials are not recommended for use in these Perchloric

Acid Hoods. Consult your Safety Officer should you have any questions.

Do not obstruct the front air foil, or rear baffle slots.

Large objects should not be placed close together and spaced above the

liner bottom to permit airflow to sweep under the equipment.

After loading the hood, wait one minute to purge airborne contaminants

from the work area.

Work Techniques

Keep all materials at least 6 inches inside of the sash, and perform all

contaminated operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.

Avoid using techniques or procedures that disrupt the airflow patterns of

the hood.

Final Purging

Upon completion of work, the hood should be allowed to operate for two to

three minutes undisturbed, to purge airborne contaminants from the work
area before shutting down blower.

Unloading Materials and Equipment

Objects in contact with contaminated material should be surface

decontaminated before removal from the hood.

All open trays or containers should be covered before being removed from

the hood.