Performance Features


Your Protector Stainless Steel Perchloric Acid Fume Hood has been engineered to
efficiently ventilate and remove Perchloric acid fumes. The Protector Perchloric
Acid Fume Hoods are designed to provide a flow of air sufficient to extract the
Perchloric acid fumes from within the work area and keep them away from the
operator to avoid ingestion, inhalation, or skin contact.

The hood interior is constructed of seamless, ground and polished 316 stainless
steel with coved interior corners. The hood also features an integral washdown
system, which helps prevent the accumulation of perchlorates. The smooth interior
finish compliments the washdown system as there are no screws or hollow
fasteners present that could harbor undesirable residue.

The Protector Perchloric Acid Fume Hood has been engineered to allow you, the
customer, the maximum work area and personnel safety. The by-pass air
configuration on the Protector Perchloric Acid Fume Hoods is totally dependent on
air being supplied to the hood from its surrounding environment for proper
operation.

Usage of materials other than Perchloric acids is not recommended. Perchloric
acid is a strong oxidizing agent, and should not be used in combination with any
organic material as the possibility of a hazardous chemical reaction or explosion
could develop.

The integral washdown system in the Perchloric Acid Fume Hood allows the
operator the ability to wash behind the baffle and ensure that a build-up of
Perchloric salts or acids does not develop in that location. All of the washdown
water from the internal spray nozzles is contained behind the baffle and does not