Chapter 4: Performance Features & Safety Precautions

Product Service 1-800-522-7658

23

Original instructions

interfere with normal operations being performed inside the hood. The washdown
water is collected in the hood drain trough.

The hood liner must be washed down manually by the operator. The internal
washdown system can be connected to the wash rings used in your hood
exhaust system. The water used in the spray wash rings is collected in the
hood drain trough once it has been allowed to flow down and cleanse the duct
thoroughly.

WARNING: It is recommended that the Perchloric Acid Hood
be washed down thoroughly after every usage for 10 minutes, at
least once a day
.

AVERTISSEMENT: Il est recommandé que le capot Perchloric
Acid être lavé soigneusement après chaque usage pendant 10
minutes, au moins une fois par jour.


1. Unique sash provides maximum visibility of 37.5" high while

conserving energy by limiting sash travel to 28". Vertical-rising sash
may be raised from a closed to 28" operating height. Exhaust volume and
blower sizing are based on the 28" height. Optional sash stops are
available to limit sash height and reduce energy usage. The Perchloric
Acid Hood has a glass sash specifically used for Perchloric Acid.

2. By-pass airflow design ensures relatively stable face velocities.

3. Large usable interior work depth and interior height of 48" provides

ample working space.

4. Baffle directs airflow to the rear of the interior to provide efficient airflow.

The baffle may be removed for cleaning purposes only.

5. Exterior access cover plates are removable for easy access to plumbing

valves when access through the sides is not available.

6. Lift-Away™ front panel provides easy access to electrical wiring, sash

weights, and lighting fixtures.

7. Energy efficient fluorescent lighting is located behind a laminated safety

glass shield mounted to the top of the hood. The factory-wired lighting is
serviceable from outside the hood cavity.

8. Low-mounted, factory-wired light and blower switches are ADA

compliant.

9. Clean-Sweep™ air foil allows air to sweep the work surface for maximum

containment.

10. Streamlined corner posts provide maximum visibility and the flexibility

to add services; services are most easily added at the factory, because of the
PVC liner.

11. All hoods are factory-prepared for up to 7 service fixtures. Additional

fixtures are available only as a factory special. Liner hole drilling is
required to add another fixture in the field.

!