12 rs232 commands, 1 output formats, 13 printing data – Ohaus CD-11 Indicator Manual en Manuel d'utilisation
Page 25: Operation (cont.) cd-11 indicator

25
4.13 Printing Data
Printing data to an external computer or printer requires that the communications parameters in the Print menu, be
set first. Refer to paragraph 2.7.6 Print Menu for proper set up.
• To print data, press Print/Units button with a short press. The display acknowledges by momentarily blinking off.
NOTE: If you hold this button down too long, the display will advance to another measuring unit.
4.12 RS232 Commands
All communication is accomplished using standard ASCII format. Characters shown in the following table are ac-
knowledged by the Indicator. Invalid command response "ES" error indicates the Indicator has not recognized the
command. Commands sent to the Indicator must be terminated with a carriage return (CR) or carriage return-line line
feed (CRLF). Data output by the Indicator is always terminated with a carriage return - line feed (CRLF).
4.12.1 Output Formats
Data output can be initiated in one of two ways; 1) By pressing Print/Units button, 2) Sending a print command (“P”)
from a computer or a PLC.
The output format is illustrated in the RS232 command table which follows.
Command
Character
Description
?
Print current mode: g, kg, lb
0S
Send weight (stable or unstable)
1S
Send stable weight only
AS
Auto send data when stable after motion
Z
Same as zero key. When scale stable, new zero established OK! response on PC screen.
When unstable, zero was not established, no response.
P
Print display data.
T
Same effect as pressing Tare Key. When scale stable, tare was established or cleared -
response OK. When unstable, "Cannot Tare" on display.
RS232 COMMAND TABLE)
4. OPERATION (Cont.)
CD-11 Indicator