2.7 Initial Setup (Cont.)
2.7.5 Readout Menu

The Readout menu is used to adapt the Indicator to environmental conditions, set measuring units on/off, parts
counting, auto zero tracking, timer on/off and retain zero data. It contains 10 submenus. Review all of the settings
available before proceeding.

Procedure

To select any of the items in the Readout menu, proceed
as follows: NOTE: If you have entered from the
preceeding menu, disregard the first step.

• With the Indicator ON, press and hold the Zero/Menu

button until MENU is displayed. When you release
the Zero/Menu button, CAL is displayed, then press
Print/Units button, until rEAd is displayed.

• Press Zero/Menu button, rESETn is displayed. This

allows resetting the readout menu to factory defaults.
rESETn = no does not reset settings. rESETy= yes
will reset the entire readout menu as follows: AL Lo,
StAb 0.5, UnOff g, Un On kg, Un On Lb, AZt 0.5, Aot
Off, rZd Off.

• Press Print/Units button,and select N or Y.

AVERAGING LEVEL
Averaging level compensates for vibration or excessive
air currents on the scale base. During operation, the
indicator continually takes weight readings from the load
cell. Successive readings are then digitally processed to
achieve a stabilized display. Using this feature, you
specify how much processing you need.

HI and LO settings are available.

HI setting:
More processing, greater stability and slower
stabilization time.

LO setting:
Less processing, less stability and faster
stabilization time.

• Press Zero/Menu button, AL LO is displayed. This is

averaging level settings. Selections are:

'Lo' - Averaging level is low
'Hi' - Averaging level is high.

• Press Print/Units button,and select LO or HI.

STABILITY
The stability range specifies the weighing results and
must be within a preset tolerance limit for a certain time to
turn the stability indicator ON. When a displayed weight
changes beyond the allowable range, the stability indica-
tor turns OFF, indicating an unstable condition. Factory
default setting is shown in bold type.

.5d Smallest range: stability indicator is ON

only when displayed weight is within .5
divisions.

1d

Normal

3d

More stable course