2.7 Initial Scale Setup (Cont.)
2.7.7 Lockout Switch Menu

Lockout Switch menu (LOCSW) is a software controlled option which can lock the settings in the Calibration, Setup,
Readout, and Print menus to prevent tampering. When used in conjunction with the Lock Switch (jumper) on the
printed circuit board, the Calibration, Setup, Readout and Print menus can be read only and not changed by an
operator.

Procedure

To select any of the items in the Lockswitch menu, pro-
ceed as follows: NOTE: If you have entered from the
preceeding menu, disregard the first step.

• With the Indicator ON, press and hold the Zero/Menu

button until MENU is displayed. When you release
the Zero/Menu button, CAL is displayed, then press
Print/Units button, until LOCSW is displayed.

• Press Zero/Menu button, LSTOFF is displayed. This

permits locking the Setup menu. OFF is unlocked, ON
is read only (locked). This menu is hidden if CAL
jumper is off.

• Press Print/Units button, and select ON or OFF.

• Press Zero/Menu button, LrdOFF displayed. This

permits locking the Readout menu. OFF is unlocked,
ON is read only (locked).

• Press Print/Units button, and select ON or OFF.

• Press Zero/Menu button, LPtOFF is displayed. This

permits locking the Print menu. OFF is unlocked, ON
is read only (locked).

• Press Print/Units button, and select ON or OFF.

• Press Zero/Menu button, LCLOFF is displayed. This

permits locking the Calibration menu. OFF is un
locked, ON is read only (locked). This menu is hid-
den if the CAL jumper is off.

• Press Print/Units button, and select ON or OFF.

• Press Zero/Menu button to end this block, END is

displayed.

• Press Zero/Menu button, Quit is displayed.

• Press Print/Units button to go to CAL or press Zero/

Menu button, Indicator returns to a weighing mode.

NOTE: At this point, the Indicator must be calibrated and
the jumper removed from the CAL connector in order to
lock out the menus. The top cover of the Indicator should
be free to gain access to the CAL jumper.