2.7 Initial Setup (Cont.)
2.7.5 Readout Menu (Cont.)
Procedure
STABILITY (Cont.)
• Press Zero/Menu button, StAb1 is displayed.The
stability range specifies the weighing results and must
be within a preset tolerance limit for a certain time to
turn the stability indicator ON. When a displayed
weight changes beyond the allowable range, the stabil-
ity indicator turns OFF, indicating an unstable condition.
0.5 d smallest range, stability indicator is ON only
when displayed weight is stable within 0.5 divisions.
1 d- stable within 1 division. 3 d-largest range, stabil-
ity indicator is ON even though displayed weight
changes 3 divisions. Factory default setting is 1.
• Press Print/Units button,and select 0.5, 1, or 3. Normal
1d stability is default/recommended.
UNITS SELECTION
• Press Zero/Menu button, Un OFF g is displayed. This
displays only when graduation is preset in grams.
• Press Print/Units button, and select ON or OFF. OFF
is the default setting.
• Press Zero/Menu button, Un ON lb is displayed. This
is unit pounds which can be turned ON or OFF. This
will be displayed when CAL UNIT kg was selected.
When lb was selected as calibration unit, kg will
display.
• Press Print/Units button,and select ON or OFF. ON
is the default setting.
• Press Zero/Menu button, Un (Pcs) OFF is displayed.
This is the Parts Counting function which can be turned
ON or OFF. Default setting is OFF.
• Press Print/Units button,and select ON or OFF.
AUTO ZERO
• Press Zero/Menu button, AZt 0.5 is displayed. This is
the Auto Zero Threshold setting. Auto Zero minimizes
the effects of temperature changes and small
disturbances on the zero reading. The Indicator
maintains the zero display until the threshold is ex-
ceeded. Settings are shown as follows:
0.5d Sets threshold to 0.5 divisions.
1d
Sets threshold to 1 division.
3d
Sets threshold to 3 divisions.
Factory default setting is 0.5d.
• Press Print/Units button,and select 0.5, 1 or 3.