2.7 Initial Setup (Cont.)
2.7.3 Load Cell Capacity Information

Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the
indicator are compatible with the scale base. Below is a Load Cell Scale Capacity (lb or kg) table. Use this table to
determine the settings of the Indicator based on the capacity and resolution of the scale base.

LOAD CELL SCALE CAPACITIES (LB OR Kg)

Grad

1000d

1200d

1500d

2000d

2500d

3000d

4000d

5000d

6000d

7500d

10000d 20000d

0.001

-

-

-

-

-

-

-

5

-

-

10

20

0.002

-

-

-

-

-

-

-

10

-

-

20

40

0.005

5

-

-

10

-

-

-

20

25

30

50

100

0.005

5

6

7.5

10

12.5

15 20 25 30 37.5 50 100

0.01 10

-

-

20

25

30

40

50

60

75

100

200

0.02 20

-

30

40

50

60

-

100

120

150

200

400

0.05 50

60

75

100

-

150

200

250

300

-

500

1000

0.1

100

120

150

200

250

300

400

500

600

750

1000

2000

0.2

200

-

300

400

500

600

-

1000

-

1500

2000

-

0.5

500

600

750 1000

-

1500

2000

2500

3000

-

5000

10000

1

1000

1200

1500 2000

2500

3000

4000

5000

-

7500

10000 20000

2

2000

-

3000

-

5000

-

-

10000

-

-

20000

-
5

5000

-

7500 10000

-

-

20000

-

-

-

-

-