3 load cell capacity information, Cd-11 indicator – Ohaus CD-11 Indicator Manual en Manuel d'utilisation
Page 11

11
2.7 Initial Setup (Cont.)
2.7.3 Load Cell Capacity Information
Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the
indicator are compatible with the scale base. Below is a Load Cell Scale Capacity (lb or kg) table. Use this table to
determine the settings of the Indicator based on the capacity and resolution of the scale base.
LOAD CELL SCALE CAPACITIES (LB OR Kg)
Grad
1000d
1200d
1500d
2000d
2500d
3000d
4000d
5000d
6000d
7500d
10000d 20000d
0.001
-
-
-
-
-
-
-
5
-
-
10
20
0.002
-
-
-
-
-
-
-
10
-
-
20
40
0.005
5
-
-
10
-
-
-
20
25
30
50
100
0.005
5
6
7.5
10
12.5
15 20 25 30 37.5 50 100
0.01 10
-
-
20
25
30
40
50
60
75
100
200
0.02 20
-
30
40
50
60
-
100
120
150
200
400
0.05 50
60
75
100
-
150
200
250
300
-
500
1000
0.1
100
120
150
200
250
300
400
500
600
750
1000
2000
0.2
200
-
300
400
500
600
-
1000
-
1500
2000
-
0.5
500
600
750 1000
-
1500
2000
2500
3000
-
5000
10000
1
1000
1200
1500 2000
2500
3000
4000
5000
-
7500
10000 20000
2
2000
-
3000
-
5000
-
-
10000
-
-
20000
-
5
5000
-
7500 10000
-
-
20000
-
-
-
-
-
CD-11 Indicator