App.1-5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 Manuel d'utilisation
Page 44

Annexe 1 Tableau de réglage utilisateur PH/ORP>
App.1-5
IM 12A01A02-12F-E
Paramètre
Valeurs par
défaut
Valeurs applicables Réglages *Réf.
Advanced
setup
Settings
No action
No action, Load
factory settings, Save
user settings, Load
user settings
5.6.1
Tag
FLXA21-PH
(Max. 12 characters)
5.6.2
Passwords Execute:
(blank)
(Max. 8 characters)
5.6.3
Commissioning:
(blank)
(Max. 8 characters)
Date/Time
Format
YYYY/MM/
DD
YYYY/MM/DD,
MM/DD/YYYY,
DD/MM/YYYY
5.6.4
Date
—
—
—
Time
—
—
—
Communication
HART
None, HART, PH201G
5.6.5
HART
Network address
0
0 à 15
PV
(pH1)
—
—
SV
Temperature1 pH1 (2), Tempera-
ture1 (2), Ref.imped-
ance1 (2), ORP1 (2),
rH1 (2), Calculated,
Redundant, Empty
TV
Empty, pH2
FV
Empty,
Temperature2
PH201G
Hold contact:
Enabled
Disabled, Enabled
Fail contact:
Fail+Warn
Fail+Warn, Fail only,
Disabled
Wash contact
Enabled
Disabled, Enabled
Interval time
6.0 h.
0.1 à 36.0 h.
Wash (Meas-
ure) time
0.5 min.
0.1 à 10.0 min.
Recovery time
0.5 min.
0.1 à 10.0 min.
Manual Wash
Enabled
Disabled, Enabled
Imp2 Wash
Disabled
Disabled, Enabled
Continuous
Wash
Disabled
Disabled, Enabled
Factory setup
—
—
—
5.6.6
PH