32
W
ILL
CLEAR
WHEN
:
•
the low pressure condition is fixed
•
treatment is stopped by pressing the
Start/Stop key.
P
OWER
FAIL
ALARM
T
RIGGERED
BY
:
•
power failure
•
machine is disconnected or switched off while delivering treatment.
I
F
A
POWER
FAIL
ALARM
SOUNDS
:
•
The flow generator stops delivering air pressure.
W
ILL
STOP
:
•
when the
Alarm Mute key is pressed, or
•
after 2 minutes, or
•
when power is restored.
!
CAUTION
Remove the mask from your face if the power fails.
H
IGH
PRESSURE
ALARM
T
RIGGERED
BY
:
•
the mask pressure exceeds the set trigger level (25 cmH
2
O) for more than 700
milliseconds.
I
F
A
H
IGH
PRESSURE
ALARM
SOUNDS
:
1. The treatment will stop.
2. Turn power off.
3. Check that the air tube and sensor tubes are connected properly.
4. Turn power back on.
5. Remove mask and perform Air Circuit Learn function.
6. Try using the flow generator one more time.
7. If the high pressure alarm activates repeatedly, discontinue use and return to
ResMed for servicing. If the alarm does not recur, then continue to use as normal.
N
OTE
The alarm system is designed to ignore coughing. However, if a cough is especially intense
and prolonged, it may trigger the high pressure alarm.
W
ILL
STOP
WHEN
:
•
the A
UTO
S
ET
CS2 is turned off.
H
IGH
LEAK
ALARM
T
RIGGERED
BY
:
•
high mask leak (greater than 30 L/min) for more than 20 seconds.
I
F
A
HIGH
LEAK
ALARM
SOUNDS
:
•
Adjust the mask to minimise leak.