ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation
Page 19

15
T
HE
A
UTO
S
ET
CS2 S
YSTEM
E
NG
LI
SH
H
UMID
A
IRE
You will need a Medium size (52cm) air tube to connect the A
UTO
S
ET
CS2 unit to the
H
UMID
A
IRE
™ (see Figure 7).
1
Make sure both the H
UMID
A
IRE
and the A
UTO
S
ET
CS2 are turned off.
Fill the H
UMID
A
IRE
with water as described in the humidifier manual. Place the filled
water chamber inside the H
UMID
A
IRE
.
2
Gently detach enough of the tube clips to release the end of the pressure sensor tube
from the A
UTO
S
ET
CS2 air tubing.
Connect the A
UTO
S
ET
CS2 air tubing to the left connector port on the humidifier, and
the medium (52cm) air tubing to the right connector port. Close the H
UMID
A
IRE
lid.
3
Place the A
UTO
S
ET
CS2 on top of the H
UMID
A
IRE
. Do not place the A
UTO
S
ET
CS2
unit underneath the humidifier. (This is to avoid water spilling into the unit.)
HumidAire
Medium air tubing
AutoSet CS2 air tubing
Pressure sensor tube
eng.book Page 15 Friday, November 3, 2006 2:43 PM