ResMed Flow Generator AutoSet CS 2 Manuel d'utilisation
Page 30

26
S
ETTINGS
MENU
You will only see the options relevant to your therapy. In CPAP (Continuous Positive
Airway Pressure) mode, the clinician sets a pressure that will be constant throughout
treatment. In ASV-CS mode, A
UTO
S
ET
CS2 continually measures your breathing and
regulates the air pressure accordingly.
M
ASK
Select the type of mask that you will be using (you will only need to adjust this if you
have more than one mask). The following table shows the setting that should be
selected for each mask type. If your mask type is not mentioned below, please contact
your clinician. Not all mask types can be used with the A
UTO
S
ET
CS2.
Table 2: Selecting mask options in the Settings menu
L
EARN
C
IRCUIT
The Air Circuit Learn function takes approximately 20 seconds to run. It allows the unit
to ’learn’ what pressure it needs to build in to its calculations, based on the
components in the system. Select LEARN CIRCUIT every time you add or remove a
component (eg mask, humidifier) or move to a significantly different altitude.
To run this function:
1. Set up the entire A
UTO
S
ET
CS2 system (flow generator, mask, humidifier etc) as
you wish to use it.
2. Make sure that the mask is unobstructed so air can flow from the mask to the flow
generator.
3. Turn on the A
UTO
S
ET
CS2 at the power switch.
4. Move through the menus until you reach LEARN CIRCUIT and select ’yes’ by
pressing the
Left key.
5. A message will remind you to check that the airflow from the mask is clear. Select
’start’ with the
Left key.
6. A progress screen is displayed while the Air Circuit Learn function is running.
Option in Settings menu
Use setting with:
Vista
V
ISTA
™
Ultra
U
LTRA
M
IRAGE
™
Full Face
M
IRAGE
™ F
ULL
F
ACE
M
ASK
S
ERIES
II
U
LTRA
M
IRAGE
F
ULL
F
ACE
M
ASK
Activa
A
CTIVA
™
Papillon
Papillon
LEARNING CIRCUIT
eng.book Page 26 Friday, November 3, 2006 2:43 PM