Labconco Protector XStream Laboratory Hoods 11081 Series Manuel d'utilisation

Page 25

Advertising
background image

Chapter 4: Performance Features & Safety Precautions

Product Service 1-800-522-7658

21

Original instructions

15. Accessory Guardian™ Digital Airflow Monitor or Guardian Airflow

Monitor continuously monitors face velocity. An audio/visual alarm
alerts the user to low airflow conditions. The right corner post is factory
prepared to accommodate the Guardian Monitor (sold separately).

16. Optional Sash Models. Sashes are offered in smooth anti-racking cable,

chain, auto-return, or Intelli-Sash.

17. Frame of Epoxy-Coated Steel and Aluminum is durable and corrosion

resistant.

18. Exhaust Connection. The hood features 12" (12.75" OD pipe) exhaust

connections sized to allow for a minimum static pressure loss through the
hood structure while providing a good transport velocity through the
exhaust system.

19. Spillstopper™ Solid Epoxy Work Surface is dished to contain spills.

Hoods with Eco-Foil utilize a large 1.0" radiused leading edge to promote
aerodynamic airflow. (Work surface is sold separately).

20. Optional Ceiling Enclosure Kits are available for a decorative facade

between the hood and the ceiling.

21. Sash Stops located 18" off the work surface provide a means of

controlling the operating height of the sash, and further reducing energy
requirements if so desired.


6

Advertising