Chapter 2: Prerequisites

Product Service 1-800-522-7658

7

Original instructions


Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a remote blower system.

Electrical Requirements


The Protector XStream Hood models feature internal wiring for the fluorescent
light assembly and light switch. All internal wiring is terminated at the single
point wiring junction box for hook-up by a qualified electrician. The blower
switch, and light switch wires are also terminated at the single point wiring
junction box for hook-up by a qualified electrician. Refer to Chapter 3: Getting
Started
and Appendix C: Protector XStream Specifications for the wiring diagram
for proper electrical installation.

Les modèles Protège-capot disposent câblage interne de l'assemblage de lumière
fluorescente et interrupteur de lumière. Tout le câblage interne est terminé à la
boîte de jonction point de câblage unique pour le raccordement par un électricien
qualifié. Le bouton du ventilateur, et fils de l'interrupteur de lumière sont
également mis fin à la boîte de jonction point de câblage unique pour le
raccordement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en
route et à l'Annexe C: Spécifications Premier Protector pour le schéma de câblage
pour l'installation électrique correcte.

Airflows and Static Pressure

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at 62.5% Static Pressure (inches of water)

Static Pressure (inches of water)

Full open

open

4'

5'

6'

8'

4'

5'

6'

8'

at 28"

at 18"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XSTREAM

100

160

705, 0.26"

930, 0.32" 1150, 0.41" 1600, 0.29"

725, 0.28"

960, 0.34" 1190, 0.44" 1655, 0.31"

80

128

565, 0.17"

745, 0.20"

920, 0.26"

1280, 0.19"

580, 0.18"

770, 0.22"

950, 0.28" 1325, 0.20"

60

96

425, 0.09"

560, 0.12"

690, 0.15"

960, 0.10"

435, 0.10"

575, 0.12"

715, 0.16"

995, 0.11"

50

80

350, 0.06"

465, 0.08"

575, 0.10"

800, 0.07"

365, 0.07"

480, 0.08"

595, 0.11"

830, 0.08"

40

64

280, 0.04"

370, 0.05"

460, 0.07"

640, 0.05"

290, 0.05"

385, 0.06"

475, 0.07"

660, 0.05"

Eco Energy Saving Air Foil

Flush Foil

Hood Type

Face

Velocity (fpm) Airflow Volumetric Rate (CFM) @

Airflow Volumetric Rate (CFM) @

Sash at

Sash at

Static Pressure (inches of water)

Static Pressure (inches of water)

62.50%

Full open

4'

5'

6'

8'

4'

5'

6'

8'

open at 18"

at 28"

Hood

Hood

Hood

Hood

Hood

Hood

Hood

Hood

XSTREAM

100

62

440, 0.10"

580, 0.12"

720, 0.16"

1000, 0.11"

455, 0.11"

600, 0.13"

745, 0.17" 1035, 0.12"

80

50

350, 0.06"

465, 0.08"

575, 0.10"

800, 0.07"

365, 0.07"

480, 0.08"

595, 0.11"

830, 0.08"

60

38

265, 0.04"

350, 0.05"

430, 0.06"

600, 0.04"

275, 0.04"

360, 0.05"

445, 0.06"

620, 0.04"

Advertising