Chapter 3: Getting Started

Product Service 1-800-522-7658

17

Original instructions

CAUTION: Natural gas should be used only in the service
fixture that has been pre-plumbed with brass tubing. Sulfur
content of the gas could cause deterioration of standard copper
supply lines.


PRUDENCE : Le gaz naturel devrait être seulement utilisé
dans l'accessoire de service qui a été pré soudé avec des tuyaux
de cuivre. Le contenu soufré du gaz pourrait causer la
détérioration des lignes d’alimentation en cuivre standard.

Sealing the Protector Hood to the Work

Surface


When the hood has been set in place, ducted, wired, and plumbed, it should be
sealed at the work surface to prevent spilled materials from collecting under the
walls of the hood. Materials such as silicone sealants are recommended to seal
the hood structure.

Certifying the Protector Fume Hood


The combination of your laboratory hood, exhaust ductwork, and exhaust blower
gives you the flexibility to change the airflow at the sash opening of your hood.
To determine the actual face velocity at the sash opening, airflow velocity
readings will need to be taken. This should be done across the sash opening of
the hood in accordance with the Industrial Ventilation Manual section on
laboratory hoods (see Appendix E: References). Labconco recommends an
average face velocity at the sash opening of 50 to 100 feet per minute for
Protector XStream high performance fume hoods. Consult Chapter 2 for proper
airflow volumes for your particular model. To ensure the performance at 50-60
feet per minute, Labconco researchers successfully challenged the Protector
XStream at face velocities lower than 50 feet per minute, under various adverse
conditions.

Your Protector Fume Hood has been tested at the factory per ASHRAE 110-1995.
All hoods achieve an “as manufactured rating” of less than 0.05 part per million
(ppm) at 4 liters per minute (lpm); AM <0.05 (consult Labconco for individual
fume hood ratings). For “field use” ASHRAE testing contact Labconco Sales
Engineering Team or Customer Service for a certified on-site contractor.

!