Appendix C: WaterPro RO System Specifications

39

Original Instructions

Turbidity:

< 1 NTU

Langlier Saturation Index*:

Negative (if positive, requires water softener
pretreatment)



*LSI is indicated in the Water Profile water analysis test treatment. Recommendations will be
provided by Labconco.

Maximum Ion Concentrations:

Iron (Total):

0.1 ppm (mg/L)

Manganese 0.1

ppm

Free Chlorine:

0.5 ppm



Environmental Conditions:

This equipment is designed to be safe under the following conditions:

• Indoor use.
• Altitude up to 6562 Ft. (2000 m).
• Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing

linearly to 50% relative humidity at 104°F (40°C).

• Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
• Transient over-voltages according to Installation Categories II (Over-Voltage

Categories per IEC 1010).

• Pollution degrees 2 normally only non-conductive foreign matter, solid, liquid, or

gaseous (ionized gases), that may produce a reduction of dielectric strength or surface
resistivity occurs. Occasionally, however, a temporary conductivity caused by
condensation must be expected, in accordance with IEC 664.