Refer to the following if your WaterPro RO System fails to operate properly. If the suggested
corrective actions do not solve your problem, contact Labconco for additional assistance.

PROBLEM CAUSES CORRECTIVE

ACTION

Unit inoperative no
display

Unit not plugged into outlet

Plug WaterPro RO into
appropriate electrical service

Circuit breaker tripped/fuse blown

Reset circuit breaker/replace
fuse

Power Switch is in the OFF position Press power button to switch

system on

Defective circuit board

Replace circuit board

Pump does not
operate and no water
is dispensed

Customer supplied
Feed valve is closed

Open the valve

Feed-water line is restricted, or no
supply

Inspect and adjust feed-water
line as required
Check inline strainer for
obstruction

Pump is defective

Replace pump