Chapter 6: Troubleshooting
33
Original Instructions
PROBLEM CAUSES CORRECTIVE
ACTION
Increase in µS
(microsiemen)
reading
High inlet TDS
This is normal. With
increased inlet TDS, RO µS
increases
RO membrane scaling or other
deposits
Clean membrane per
instructions in RO cleaning
section
Excessively high µS
RO membrane fouled
Biofilm formed
Clean RO membrane per
instructions in RO cleaning
section
RO poisoned
Chlorine damage
Replace RO membrane and
prefilters per installation
instructions