Chapter 6: troubleshooting – Labconco WaterPr PS Systems 90007 Series Manuel d'utilisation
Page 34

Chapter 6: Troubleshooting
30
Original Instructions
PROBLEM
CAUSES
CORRECTIVE
ACTION
Recirculation
pump operates but
no water is
dispensed (cont.)
Return line check valve is
defective
Replace check valve assembly.
Recirculation pump is defective
Replace recirculation pump.
Reduced flow at
dispense valve/gun
Optional final filter is blocked
Replace final filter.
Restricted tube in the feed line or
polishing system
Inspect polishing loop tubing for
any restrictions.
Water quality
display acting
erratically
New cartridges installed in
system
Purge all of the air out of the
polishing system.
Air trapped in resistivity cell
Purge all of the air out of the
polishing system.
Deionization
cartridges
are
exhausted
Replace
the
deionization
cartridges.
Polishing loop resistivity cell not
connected to wiring harness
Reconnect cell to wiring harness.
Polishing loop resistivity cell is
defective
Replace
cell.
Polished water
will not rinse up to
desired quality
Cartridges are past expiration
date
The shelf life of unopened
cartridges is two years from the
date of manufacture. Replace the
expired cartridges with new ones.
Check the manufactured date.
Cartridges installed in wrong
sequence
Install the cartridges as described
in the “Initial Operation and
Cartridge Replacement” section.
Water is bypassing the cartridges Ensure that the O-rings are at the
top of the deionization cartridges
and the posts at the top of the
bowls are inserted correctly.