Chapter 3: Getting Started
15
Original Instructions
Cartridge Installation
1. Remove the four polypropylene sump bowls by placing the spanner wrench as high as
possible around the bowls. Unscrew each bowl by turning clockwise. With a twisting
motion, install the carbon filter in the first bowl and other cartridges as indicated in the
table below. Check each cartridge to ensure the two black O-rings are seated properly into
the sump top. Wetting of these O-rings will aid in installation. Carefully thread the bowls
back into the position and hand tighten only to seal the bowl O-rings.
CARTRIDGE INSTALLATION SEQUENCE
(BOWLS ARE NUMBERED 1-4 FROM LEFT TO RIGHT)
HAND TIGHTEN SUMPS ONLY - DO NOT USE WRENCH
Model
Bowl
Number
Cartridge Type
Kit Number
Catalog
Number
General Chemistry
Models
90005 Series
1
2
3
4
Carbon
Deionization
Deionization
Deionization
9047101
9007201
9007301
9007301
9007301
HPLC and
HPLC/UF Hybrid
90006 Series
1
2
3
4
Carbon
Deionization
Deionization
Organic Adsorption
9047201
9007201
9007301
9007301
9053300
Life Science Models
90007 Series
1
2
3
4
Carbon
Deionization
Deionization
Deionization
9047401
9007201
9007301
9007301
9007301
Ultrafilter is included
and installed on all UF
models
9104400
Hollow Fiber Final
Filter (if required)
9092900
2. Place a suitable container or a hose that goes to a drain over the dispense valve or dispense
gun nozzle and SLOWLY open the inlet supply valve.
3. Plug the unit into an electrical outlet and turn the power switch on.
NOTE: When activating a new Polishing System, or one that has new cartridges installed, the
system will initially be full of air. This may cause the water quality display to operate
erratically or incorrectly. This is common, and the display will resume normal operation when
all of the air has been displaced from the system. Allow system to dispense to drain until all air
has been removed. On ultrafiltered models, turn unit off and press dispense. Repeat until all air
has been removed from the system. This is a normal air purging procedure.