ZOLL AED Pro Rev G Manuel d'utilisation
Page 102

A
NNEXE
A
S
PÉCIFICATIONS
A–16
www.zoll.com
9650-0350-02 Rév. G
Les performances relatives aux arythmies sont rapportées conformément à l’article, RE Kerber,
LB Becker, JD Bourland, RO Cummins, AP Hallstrom, MB Michos, G Nichol, JP Ornato,
WH Thies, RD White, BD Zuckerman, « Automated External Defibrillators for Public
Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis
Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety, » Circ J Am
Heart Assoc. 1997;95:1677-1682.
Références
Young KD, Lewis RJ. « What is confidence? Part 2: Detailed definition and determination
of confidence intervals ». Ann Emerg Med. September 1997;30:311-318.
Beyer WH. Percentage Points, F-Distribution Table. CRC Standard Mathematical Tables.
28
th
ed. Boca Raton, Fla: CRC Press; 1981:573.
Tableau A-5.
Résultats relatifs aux performances cliniques (patients enfants)
Rythmes
Taille de
l’échantillon
(sur 9 secondes)
Objectifs de
performances
Performances
observées
Limite inférieure
de l’intervalle
de confiance à
90 % unilatéral
Choquable
(49 patients)
Sensibilité
FV brute
42
>90 %
100 % (42/42)
93,1 %
TV rapide
82
>75 %
93,9 % (77/82)
87,6 %
Non choquable
(155 patients)
Spécificité
RSN
208
>99 %
100 %
(208/208)
98,6 %
FA, BS, TSV
a
,
bloc cardiaque
idioventriculaire,
ESV
a. 161 enregistrements de rythmes anormaux sur 348 concernaient des TSV (72 patients).
Les fréquences cardiaques en cas de TSV sont comprises entre 152 et 302 battements
par minute.
348
>95 %
99,4 %
(346/348)
98,2 %
Asystolie
29
>95 %
100 % (29/29)
90,2 %
Intermédiaire
(16 patients)
FV fine
0
Uniquement
rapportées
—
—
Autres TV
40
Uniquement
rapportées
90 % (36/40)
78,6 %