Before you install your blower, you need to prepare your site for installation. A
dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
The location requirements for your installation site.
The mounting support requirements for your installation site.
The electrical power requirements for your installation site.
The fume hood controls, controls wiring, and controls modifications for
your installation site.
The space requirements for your installation site.
Refer to Appendix B: Blower Dimensions for complete blower dimensions.
Refer to Appendix C: Blower Environmental Conditions for complete
environmental conditions and wiring diagram.
Location Requirements
In positioning your blower, care should be taken to make sure that it is away from
all other types of air handling equipment (intake fans, air conditioning units, etc.).
Your blower should also include a minimum of 7 ft. of ducting above the roofline
to generate proper air dispersion of materials being exhausted through the blower.
Also consider the location of the blower inlet with respect to the hood you are
exhausting. Proper planning and layout are essential in selecting a blower
location.
- Intelli-Sense Multi-Speed Blowers 7183412 Intelli-Sense Multi-Speed Blowers 7183411 Intelli-Sense Multi-Speed Blowers 7183410 Intelli-Sense Multi-Speed Blowers 7181813 Intelli-Sense Multi-Speed Blowers 7181812 Intelli-Sense Multi-Speed Blowers 7181811 Intelli-Sense Multi-Speed Blowers 7181810 Intelli-Sense Multi-Speed Blowers 7061113 Intelli-Sense Multi-Speed Blowers 7061112 Intelli-Sense Multi-Speed Blowers 7061111 Intelli-Sense Multi-Speed Blowers 7061110