Chapter 4: Performance Data and Safety Precautions
Product Service 1-800-522-7658
24
Original instructions
Speed
(RPM)
Optimum Airflow (CFM) and Static Pressure (inches of water)
1800 MAX
750 CFM @ 1.7" to 1025 CFM @ 1.55"
1500
600 CFM @ 1.2" to 1050 CFM @ 0.5"
1250
500 CFM @ 0.8" to 850 CFM @ 0.37"
1000
400 CFM @ 0.6" to 700 CFM @ 0.25"
750
300 CFM @ 0.3" to 475 CFM @ 0.12"
500
200 CFM @ 0.12" to 325 CFM @ 0.06"
300
175 CFM @ 0.06" to 275 CFM @ 0.04"
Practical range is 175 CFM, 0.06" to 1025 CFM, 1.5"
0
0.25
0.5
0.75
1
1.25
1.5
1.75
2
150
250
350
450
550
650
750
850
950 1050 1150 1250 1350
St
ati
c
Pressure
(in.
w
.g.)
Airflow Rate (CFM)
Models 71818-10, -11, -20, -21
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 10" FRP inlet
500
750
1000
Optimum
Blower
Performance
Note: Ambient Temperatures
Inside Motor Housing Greater
than 125 °F May Reduce
Maximum Blower Performance
1250
1500
1750
1800
SEE
NOTE
- Intelli-Sense Multi-Speed Blowers 7183412 Intelli-Sense Multi-Speed Blowers 7183411 Intelli-Sense Multi-Speed Blowers 7183410 Intelli-Sense Multi-Speed Blowers 7181813 Intelli-Sense Multi-Speed Blowers 7181812 Intelli-Sense Multi-Speed Blowers 7181811 Intelli-Sense Multi-Speed Blowers 7181810 Intelli-Sense Multi-Speed Blowers 7061113 Intelli-Sense Multi-Speed Blowers 7061112 Intelli-Sense Multi-Speed Blowers 7061111 Intelli-Sense Multi-Speed Blowers 7061110