Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
19
Original instructions
6. Energy efficient fluorescent lighting is located behind a laminated safety
glass shield mounted to the top of the hood. The factory-wired instant
start T8 lighting is serviceable from outside the hood cavity.
7. Low mounted, factory-wired light and blower switches are ADA
compliant.
8. All Air Foils allow air to sweep the work surface for maximum
containment. The Clean-Sweep™ openings create a constant barrier from
contaminants. In addition, should the operator inadvertently block the
airflow entering, the air continues to enter from under the air foil and
through the Clean-Sweep openings.
9. Streamlined corner posts provide maximum visibility and the flexibility
to add services after installation.
10. All hoods are factory prepared for up to 8 service fixtures (4 service
fixtures if Side View Window option is selected).
11. Duplex electrical receptacles are mounted on the right and left corner
posts as requested. Receptacles are factory-wired to hood single point
junction box.
12. Shipped fully assembled and eliminates the need for costly onsite
assembly.
13. Accessory Guardian™ Digital Airflow Monitor or Guardian Airflow
Monitor continuously monitors face velocity. An audio/visual alarm
alerts the user to low airflow conditions. The right corner post is factory
prepared to accommodate the Guardian Monitor (sold separately).
14. Sash Alarm. Audible alarm sounds if both sashes are raised to alert user.
15. Frame of epoxy-coated steel and aluminum is durable and corrosion
resistant.
16. Exhaust Connection. The hood features 12" (12.75" OD pipe) exhaust
connections sized to allow for a minimum static pressure loss through the
hood structure while providing a good transport velocity through the
exhaust system.
17. Spillstopper™ Solid Epoxy Work Surface is dished to contain spills.
(Work surface is sold separately).
18. Optional Ceiling Enclosure Kits are available for a decorative facade
between the hood and the ceiling.
19. Optional Sash Stops provide a means of controlling the operating height
of the sash.