Chapter 2: Prerequisites
Product Service 1-800-522-7658
8
Original instructions
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch outside diameter,
copper (brass for natural gas), and equipped with an easily accessible shut-off
valve, should disconnection be required. If the service line pressure exceeds 40
PSI, it must be equipped with a pressure regulator to reduce the line pressure.
Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in Appendix B: Protector
Pass-Through Dimensions.