4.3 Properties of the RS 232 Interface
756/831 KF Coulometer, Instructions of Use
99
Coulometer as Sender :
max. characters:
2 characters at 300...9600 baud
16 characters at
≥
19200 baud
Software-Handshake, SWline
Handshake input ports on the Coulometer (CTS) are not checked.
Handshake output ports (DTR, RTS) are set by the Coulometer.
The Coulometer has an input buffer which can accept up to 512 characters.
Coulometer as Receiver :
Coulometer
external device
Coulometer
external device
Time
Data output
Data input
RxD
TxD
XOFF
XON
L
F
max.
characters
Data output
disabled
Data output
enabled
Data output
Data input
max. 128
characters
384 characters
XOFF
XON
L
F
RxD
TxD