4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
96
Silo:
The silo memory is deleted. Same function as delete entire silo.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
4.2.2.99. Setup.RamInit $G
Initializes instrument, see page 110. All parameters are set to their default value and
error messages are cleared. The user and silo memories will be deleted.
Command only possible in the inactive state of the Coulometer.
4.2.2.100. Setup.InstrNo $G
Setup.InstrNo.Value serial
number, 8 ASCII characters
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
4.2.2.101. Diagnose.Report $G
Output of the report containing the adjustment parameters. The Coulometer has to be
in its inactive basic state.
4.2.2.102. Diagnose.Simulation.Keycode
0...29
0
1
2
3
4
28
21
26
8
9
10
11
12
5
13
29
16
17
18
19
20
25
27
24
15
6
22
23
4.2.2.103. Diagnose.ScreenDump $G
The content of the 756 Screen will be dumped to the COM which is given for manual
reports (key <CONFIG>, >peripheral units). A screen dump onto the internal printer
is not possible.
4.2.2.104. Diagnose.IntPrinter.HeatTime 1...4.0...10
Diagnose.IntPrinter.MotorSpeed 2...3.0...9
Settings for the internal printer.
Entering a keycode is like pressing the corre-
sponding key. The keys have the following key-