Chapter 6: Maintaining the Protector Echo

Product Service 1-800-522-7658

84

Original instructions


Press

“X”

again and you should be back to the main menu.


You should see:

(Access Granted)

>Hood Status menu
Administration menu



Step 17 – Test the hood

Now turn on your fans and lights. If everything is connected correctly, it should
operate properly. Once you determine the fans and lights are operational, turn it
off. Wait 30 seconds for it to completely shut down.

If the control has no errors proceed to the next step.

If the control panel does have errors

Check all your connections again (make sure the wires are securely in

their slots).

Make sure you have your fans in the correct position from left to right

1,2,3,4, or 5.

Check the “Dip” switches on the front of each fan.

Based on their position from left to right:

Fan #1 should be

On, Off, Off, Off = Down, Up, Up, Up

Fan #2 should be

On, Off, Off, On = Down, Up, Up, Down

Fan #3 should be

On, Off, On, Off = Down, Up, Down, Up

Fan #4 should be

On, Off, On, On = Down, Up, Down, Down

Fan #5 should be

On, On, Off, Off = Down, Down, Up, Up


If you need to adjust the

DIP

switch, you may use a very small regular screw

driver or even a pencil. Make your adjustments by putting light pressure on the

DIP

switch to either move it up

(off)

or down

(on)

.


Reboot the hood by sliding your finger nail under the black fuse cartridge located
on the left side of the controller, slide it out to break the connection, wait 1 min
and plug it back in.

If you still have a problem, write down what the control module screen displays
and call Labconco Product Service. Do not attempt another hood until you fix the
issue.