Original instructions




C

HAPTER

6

M

AINTAINING

T

HE

P

ROTECTOR

E

CHO

Required maintenance includes routine replacement of the Neutrodine carbon
filters, HEPA filters, and fluorescent light. The security and communication
options are also listed.

Review this chapter on maintenance for the following:

Routine Maintenance

Initial Certification

Re-Certification

User Alarm Code List

Tool List

Neutrodine SDS for disposal

Partial Filter Replacement Procedure

Reconfiguration Process (listed in Chapter 3) – needed for most

maintenance procedures.

Light and/or Circuit Board Replacement

Fan and/or Circuit Board Replacement

Fan Box Replacement

Electro (Acid/Gel) Sensor Replacement – required every 2 years.

Secure Mode (see Appendix G)

gGuard Checklist and Software Option (see Appendix E)

BACnet Checklist Option (see Appendix F)

Routine Maintenance Schedule

Weekly

Wipe down the interior surfaces of the filtered hood with a suitable dish

soap, solvent, disinfectant, or cleaner, depending upon the usage of the
unit and allow to dry.

Using a damp cloth, clean the exterior surfaces of the filtered hood,

particularly the front and top to remove any accumulated dust.

Operate the fan system, noting the airflow velocity on the display and

airflow through the filtered hood using a source of visible smoke.