5.10 The remote control tree
:
.WetPart
Dosing unit definitions
├
.WetPartNo
Dosing unit no.
1...12
│ ├
∗
│
.1
Port number 1
-
│ ├
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ │ ├
.Length
Tubing length
0...1000...30000 mm
│ │ ├
└
.Diameter
Tubing diameter
0...2...20 mm
│ │
│ │
.2
Port number 2
-
│ ├
.MaxRate
Max. dosing rate
0.01…160 ml/min
│ │ ├
.Length
Tubing length
0...250...30000 mm
│ │ ├
└
.Diameter
Tubing diameter
0.1…2…20 mm
│ │
│ │
.3
Port number 3
-
│ ├
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ │ ├
.Length
Tubing length
0...1000...30000 mm
│ │ ├
.Diameter
Tubing diameter
0.1...2...20 mm
│ │ └
│ │
.4
Port number 4
│ └
.MaxRate
Max. dosing rate
0.01...160 ml/min
│ ├
.Length
Tubing length
0...1000...30000 mm
│ ├
└
.Diameter
Tubing diameter
0.1...2...20 mm
│
│
.Rsset
RS232 Settings
└
.Baud
Baud rate
300, 600, 1200,
├
2400, 4800, 9600
│
.DataBit
Number of data bits
7, 8
├
.StopBit
Number of stop bits
1, 2
├
.Parity Parity
even,
odd,
none
├
.Handsh Handshake
Hws, HWfull,
├
SWchar, SWline,
│
none
│
└
.CharSet Character
set
IBM, HP, Epson,
Seiko,
Citizen
5.10.4 &Info
Object Description
Input
range
Reference
& Root
:
Info
Current data
-
├
:
.Report
Report definition
├
└
.Select Report
type
short, full, config, param,
│
usermeth,
all
│
│
.DetermData
Determination data
└
.Mplist
Meas. point list of temperature
-
├
.1
Meas. point 1
-
:
├
└
.Y
Temperature in °C
read only
:
:
└
.400
Meas. point 400
774 Oven Sample Processor, Mode d'emploi
107