4 rotors, 1 rotor fa-45-12-17, Rotors 2.4.1 – Eppendorf Centrifuge 5427 R Manuel d'utilisation
Page 9: Rotor fa-45-12-17

Product description
Centrifuge 5427 R
English (EN)
9
2.4
Rotors
2.4.1
Rotor FA-45-12-17
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which are marked with the name of the centrifuge (e.g., 5427
R).
Rotor FA-45-12-17
Aerosol-tight fixed-angle rotor
for 12 reaction
tubes
Max.
g-force:
20
598
×
g
Max. speed:
14
000
rpm
Max. load (adapter, tube
and contents):
12
×
9.5
g
Vessels
Vessel
Adapters
Adapter base
Max. g-force:
Capacity
Order no.
(international)
Tube diameter
Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel
–
–
20
598
×
g
5
mL
Ø
17
mm
14
000
rpm
-/12
9.4
cm
Vessel
5820
768.002
open
17
530
×
g
1.5
mL/2.0
mL
Ø
11
mm
14
000
rpm
1/12
8.0
cm
HPLC vessel
5820
770.007
flat
16
215
×
g
Ø 11
mm
14
000
rpm
1/12
7.4
cm
Cryo tube
5820
769.009
flat
18
188
×
g
1.0
mL
–
2.0
mL
Ø
13
mm
14
000
rpm
1/12
8.3
cm