Eppendorf Centrifuge 5427 R Manuel d'utilisation
Page 43

Troubleshooting
Centrifuge 5427 R
English (EN)
43
Err 22
Electronics fault.
Switch centrifuge off and back on
again after > 20
s.
NO E-FAN (ERR 23)
Electronic fan is faulty.
Switch centrifuge off and back on
again after > 20
s.
Err 24
Failure at the compressor.
Allow the centrifuge to cool down and
repeat the run.
INT (ERR 25)
Power failure during a run.
Check the mains connection.
Err 28
Error during speed check.
Switch centrifuge off and back on
again after > 20
s.
Err 29
Set g-force/speed too high, e.g. after a
rotor change (see
Check the g-force/speed.
Repeat the run.
LID (ERR 30)
• Centrifuge lid cannot be locked.
• Centrifuge lid cannot be released.
Try again to close centrifuge lid.
Switch the centrifuge off and back on.
Press the
open key.
If the error occurs repeatedly:
Switch off centrifuge.
Actuate emergency release(see
Emergency release on p. 44).
OFF
• Displayed after power off.
Symptom/message Cause