6-3. Etalonnage lorsque la fonction HOLD est activée
IM 12D7C22-F-H
Calibration 6-3
µ
S / c m
HOLD
µ
S / c m
ENT
HOLD
µ
S / c m
ENT
HOLD
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
µ
S / c m
NO
YES
HOLD
YES
µ
S / c m
NO
YES
HOLD
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
µ
S / c m
NO
YES
HOLD
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
µ
S / c m
NO
YES
HOLD
µ
S / c m
NO
YES
NO
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
µ
S / c m
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
MEASURE
DISPLAY 1
HOLD
DISPLAY 2
CAL 2
CAL 1
Note: See section 4-5for full
display acces function
Appuyer sur MODE
CAL1/CAL2 apparaît YES/NO
clignote. Se SETP s’affiche,
appuyer d’abord sur NO
Plonger le capteur dans la
solution standard. Appuyer
sur YES.
Régler la valeur à l’aide des touches >, ^ et ENT
Sélectionner le digit clignotant avec > et augmenter
sa valeur à l’aide de ^.
Valider avec ENT
WAIT s’affiche brievement CAL.END apparaît
L’étalonnage est terminé. Replonger le capteur
dans le proceé. Appuyer sur YES pour repasser en
mode mesure.
HOLD s’affiche, appuyer sur NO pour revenir au
mode mesure.
Note: voir section 4-5.