2. procédure d’étalonnage – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer Manuel d'utilisation
Page 60

IM 12D7C22-F-H
6-2. Procédure d’étalonnage
6-2 Etalonnage
ENT
M
Ω .
c m
ENT
M
Ω .
c m
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
NO
YES
M
Ω .
c m
NO
YES
M
Ω .
c m
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES to return to the measuring mode.
ENT
M
Ω .
c m
M
Ω .
c m
YES
NO
YES
M
Ω .
c m
NO
NO
MODE
YES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
M
Ω .
c m
CAL 2
CAL 1
MEASURE
DISPLAY 1
HOLD
DISPLAY 2
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
Note: See section 4-5 for full
display access details
Appuyer sur MODE
CAL1/CAL2 apparaît YES/NO
clignote. Se SETP s’affiche,
appuyer d’abord sur NO
Plonger le capteur dans la
solution standard. Appuyer
sur YES.
Régler la valeur à l’aide des touches >, ^ et ENT
Sélectionner le digit clignotant avec > et augmenter
sa valeur à l’aide de ^.
Valider avec ENT
WAIT s’affiche brievement CAL.END apparaît
L’étalonnage est terminé. Replonger le capteur
dans le proceé. Appuyer sur YES pour repasser en
mode mesure.
Note: voir section 4-5.