Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity Manuel d'utilisation
Page 34

27
IM 12D08N05-01F-E
5
STRUCTURE
DES
MENUS
Menu
Paramètre
Valeurs
Etendue
par défaut
min.
max.
PID-control S1
Setpoint
250.0 µS/cm
- inf
+ inf
PID-control S1
Range
50.0 µS/cm
- inf
+ inf
PID-control S1
Manual Reset
0%
0%
100%
PID-control S1
I-time
3600 s
1 s
3600 s
PID-control S1
D-time
0 s
0 s
60 s
Duty cycle
DC period time
10 s
1 s
1800 s
Pulse freq.
Max. pulse freq.
70 p/min
1 p/m
120 p/m
mA1 (simulate)
Expire time
0.0 s
0 s
1800 s
Alarm S1
Setpoint
450.0 µS/cm (high)
- inf
+ inf
Alarm S2
Setpoint
50.0 µS/cm (low)
- inf
+ inf
Alarm S1
Hysteresis
4.500 µS/cm
0 mS/cm
+ inf
Alarm S1
Delay Time
0.2 s
0 s
+ inf
Alarm S1
Expire Time
0.0. s
0 s
1800 s
Hold
Fixed value mA1
12 mA
3.6 mA
21 mA
Hold
Fixed value mA2
12 mA
3.6 mA
21 mA
S2, S3, S4 semblable à S1
Enter
PI-control S1
Range
(proportional)
Direction
Setpoint
I-time
250.0μS/cm
50.00μS/cm
Reverse
3600 s
Alarm
Control
Hold
USP
Fail
Off
Simulate