Set receive to transmit delay, Report number of floats, Report receive to transmit delay – VEGA VEGAPULS 65 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Manuel d'utilisation
Page 66: Codes de défaut

66
11 Annexe
VEGAPULS 65 • Protocole Modbus et Levelmaster
41365-FR-121120
Set Receive to Transmit Delay
Request:
Paramètres
Length
Code/Data
Set Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 bis 250),
default = 127 ms
Response:
Paramètres
Length
Code/Data
Set Receive to Transmit Delay
6 characters ASCII
UuuROK
Report Number of Floats
Request:
Paramètres
Length
Code/Data
Set Receive to Transmit Delay
4 characters ASCII
UuuF
Response:
Paramètres
Length
Code/Data
Set Receive to Transmit Delay
5 characters ASCII
UuuFn
n = number of measurement values
(0, 1 or 2)
Report Receive to Transmit Delay
Request:
Paramètres
Length
Code/Data
Report Receive to Transmit Delay
4 characters ASCII
UuuR
Response:
Paramètres
Length
Code/Data
Report Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 bis 250),
default = 127 ms
Codes de défaut
Error Code
Name
EE-Error
Error While Storing Data in EEPROM
FR-Error
Erorr in Frame (to short, to long, wrong data)
LV-Error
Value out of limits