11 Annexe
POINTRAC 31 • Foundation Fieldbus
41779-FR-130621
Hex
Quality
Sub-Status
Condition
0x44
Uncertain
Last Usable Value
Three or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
0x48
Uncertain
Substitute
On user error codes: 29
0x4C
Uncertain
Initial Value
After startup of device or channel assignment
in AIFB was changed. The channel unit may be
unknown until next "Process Data Update"
On user error codes: 13
0x80
Good (NC)
Non-specific
No errors concerning to channel handling
On user error codes: 0, 33, 45, 71, 126, 127
Mapping of User Error Codes to FF Field Diagnostics
Priority
Description
User Error Codes
NE-107 Status
31
Hardware failure
40, 41
FAILURE
30
Memory failure
34, 35, 36, 37
FAILURE
29
Software failure
80
FAILURE
28
Paramererization corrupt
72
FAILURE
27
Undefined 27
-
-
26
Undefined 26
-
-
25
Paramererization error
16, 17, 25, 52, 53, 57, 66, 117,
120
FAILURE
24
Conflict in MGC
121, 122, 141
FAILURE
23
Communication error in MGC
8
FAILURE
22
MGC slave reports error
38
FAILURE
21
Undefined 21
-
-
20
Undefined 20
-
-
19
Undefined 19
-
-
18
Detector temperature critical
125
OUT_OF_SPEC
17
Error while auto-standardization
73
FAILURE
16
Excessive radiation
123, 124
FAILURE
15
Input out of bounds
13
OUT_OF_SPEC
14
Error while signal processing
68
FAILURE
13
Undefined 13
-
-
12
Undefined 12
-
-
11
Undefined 11
-
-
10
Undefined 10
-
-
9
Undefined 9
-
-
8
Undefined 8
-
-
7
AITB simulated
29
FUNCTION_CHECK
6
Undefined 6
-
-