10 Annexe
VEGABAR 83 • Protocole Modbus et Levelmaster
46295-FR-131120
Register Name
Register Number
Type
Note
PV
2202
DWord
Primary Variable in Byte Order BACD (Middle Endian)
SV
2204
DWord
Secondary Variable in Byte Order BACD (Middle En-
dian)
TV
2206
DWord
Third Variable in Byte Order BACD (Middle Endian)
QV
2208
DWord
Quarternary Variable in Byte Order BACD (Middle En-
dian)
Unit Codes for Register 104, 108, 112, 116
Unit Code
Measurement Unit
32
Degree Celsius
33
Degree Fahrenheit
40
US Gallon
41
Liters
42
Imperial Gallons
43
Cubic Meters
44
Feet
45
Meters
46
Barrels
47
Inches
48
Centimeters
49
Millimeters
111
Cubic Yards
112
Cubic Feet
113
Cubic Inches
10.4 Instructions RTU Modbus
FC3 Read Holding Register
Cette instruction permet de lire un nombre quelconque (1-127) de registres "holding". Le registre de
démarrage à partir duquel la lecture doit être effectuée et le nombre de registres sont transférés.
Request:
Parameter
Length
Code/Data
Function Code
1 Byte
0x03
Start Address
2 Bytes
0x0000 to 0xFFFF
Number of Registers
2 Bytes
1 to 127 (0x7D)
Response:
Paramètres
Longueur
Code/Data
Function Code
1 Byte
0x03