Maintenance and Troubleshooting
5
Display
Action
An electronic error occurred, or in rare cases, a used test strip was removed and
reinserted. Turn the meter off and on, or take the battery out for 20 seconds and
reinsert it. Perform a blood glucose or control test.
The temperature is above or below the proper range for the system. Refer to the
test strip package insert for system operating conditions. Move to an area with the
appropriate conditions, wait 5 minutes, and repeat the blood glucose or control test.
Do not artificially heat or cool the meter.
The battery is almost out of power. Change the battery now. If the message
reappears after the battery has been replaced, remove the battery again, press any
meter button, then reinsert the battery.
The time and date settings may be incorrect. Make sure the time and date are
correct and adjust, if necessary.