3.2. Remote control commands
719 S Titrino
76
3.2.2 Description of the remote control commands
3.2.2.1. Mode
$G, $S, $H, $C
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size af-
ter the start (see 3.2.2.25) as well as after inquiries of calibration temperature
and pH values of buffers (see 3.2.2.31 and 3.2.2.32).
3.2.2.2. Mode.QuickMeas $G,
$S
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the <meas/hold> key.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.
3.2.2.3. Mode.Select SET, MEAS, CAL
Mode.SETQuantity pH, U, Ipol, Upol
Mode.MEASQuantity pH, U, Ipol, Upol, T
Selection of the standard mode. Mode and the measured quantity belong to
the complete selection.
If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.
3.2.2.4. Mode.Name read
only
Name of the current method in the working memory. $Q sends 8 ASCII char-
acters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.61.
3.2.2.5. Mode.Def.Formulas.1.Formula
EPX, CXX, RSX, +, -, *, /, (, )
Mode.Def.Formulas.1.TextRS
up to 8 ASCII characters
Mode.Def.Formulas.1.Decimal 0...2...5
Mode.Def.Formulas.1.Unit
up to 6 ASCII characters
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 39f.
Example: "(EP2-EP1)*C01/C00"
In addition to the formula, a text for result output, the number of decimal
places and a unit for the result output can be selected. "No unit" is selected
with the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is out-
putted in the report full, short. It is used for the result and the corresponding
mean value.
3.2.2.6. Mode.Def.ComVar.C30
RSX, MNX, EPX, CXX
Mode.Def.ComVar.C31
etc., up to .C39